Shenzhen PM-9000 Express Betreibershandbuch Seite 251

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 339
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 250
O2 COMPEN
The options include: OFF, 30%, 40%, 50%, 60%, 70%,
80%, 90%, and 100%.
When the O2 concentration exceeds 60% and it is not being monitored, turn on this switch.
SWEEP Used to select the speed to scan the screen waveforms.
Options: 6.25mm/s and 12.5mm/s.
WORK MODE
Options: MEASURE and STANDBY.
To monitor the anesthetic gas, select the MEASURE option. Otherwise, select the
STANDBY option.
Alarm Setup Menu
Select ALARM SETUP >> in AG SETUP menu. The following menu appears.
Figure 18-5 Alarm Setup Menu
In the ALARM SETUP menu-1, you can set the following items.
ALM Alarm on/off status
ON: When the AG has an alarm, the system will give alarm
indications and save the alarm information.
OFF: The system will not give alarm indications when an
AG alarm occurs. Instead it will display
on the right of
the GAS label in the parameter window.
ALM LEV
Options: HIGH and MED.
18-7
Seitenansicht 250
1 2 ... 246 247 248 249 250 251 252 253 254 255 256 ... 338 339

Kommentare zu diesen Handbüchern

Keine Kommentare